A large-scale Indian clinical study validating V-CHEK against mercury sphygmomanometer — 250 patients at Sri Jayadeva Institute of Cardiovascular Sciences and Research, Bengaluru.
Blood pressure monitoring is undergoing a paradigm shift from traditional cuff-based methods to cuffless wearable technology.
In the era of advanced digital health technology, the approach to blood pressure (BP) monitoring is shifting from in-office and home BP monitoring toward cuffless and wearable technology. Traditional cuff-based methods — introduced by Scipione Riva-Rocci almost 125 years ago — can be cumbersome, inconvenient, and potentially inaccurate, especially for frequent measurements.
Advancements in sensor technology, signal processing, and machine learning have opened opportunities for cuffless methods that establish precise relationships between biomedical signals like photoplethysmography (PPG) and changes in BP.
Conventional BP devices using inflatable cuffs offer limited snapshots. Cuffless technologies enable continuous monitoring, potentially revolutionizing hypertension management.
Compare accuracy of V-CHEK cuffless BP monitor against mercury sphygmomanometer and pulse oximeter for BP, heart rate, and SpO₂ in resting conditions.
Cross-sectional study with 250 patients. Statistical analysis included Bland-Altman agreement assessment with 95% confidence intervals.
Strong agreement across all vital parameters measured
95% CI: −0.76 to 0.004
95% CI: 0.52 to 1.03
95% CI: −0.25 to 0.09
Pearson correlation coefficient
| Metric | Result |
|---|---|
| SBP Bias | −0.38 mmHg (95% CI: −0.76 to 0.004) — 3.6% outside limits |
| DBP Bias | 0.77 mmHg (95% CI: 0.52 to 1.03) — 4.6% outside limits |
| Pulse Rate Bias | −0.08 bpm (95% CI: −0.25 to 0.09) — 5.4% outside limits |
| SpO₂ Bias | −0.17% (95% CI: −0.24 to −0.10) — 4.3% outside limits |
| SBP Correlation | 0.959 (95% CI: 0.953–0.965) |
| DBP Correlation | 0.949 (95% CI: 0.941–0.955) |
| PR Correlation | 0.982 (95% CI: 0.979–0.985) |
| SpO₂ Correlation | 0.808 (95% CI: 0.782–0.832) |
Non-invasive, cuffless BP monitor measuring BP, heart rate, and oxygen saturation
250 patients with cardiovascular diseases in Bangalore, India
| Total Participants | 250 patients |
|---|---|
| Male / Female | 56.4% / 43.6% |
| Average Age | 51.6 ± 13.1 years |
| Inpatients | >80% of subjects |
| With Hypertension | 154 patients |
| With Diabetes | 105 patients |
| Family History of HT | 66.9% |
Inclusion Criteria:
Exclusion Criteria:
Patient rested 5 min seated, hand open and relaxed, legs uncrossed.
Three reference measurements (Ref1, Ref2, Ref3) using mercury sphygmomanometer. Average SBP and DBP entered in app.
Within 60s of reference, place finger on sensor. All four vital signs displayed in 40–60 seconds.
Published in: Journal of Indian College of Cardiology, Volume 14, Issue 3, July–September 2024
Authors: Rahul S. Patil, C. N. Manjunath, K. R. Nishanth
Institution: Sri Jayadeva Institute of Cardiovascular Sciences and Research, Bengaluru
DOI: 10.4103/jicc.jicc_46_23